Validation set of temperature and pressure recorders for testing and documenting conditions during sterilization of instruments in steam sterilizers in medical and laboratory facilities, in accordance with ISO 17665
The validation kit allows the following equipment to be verified:
- Healthcare:
- Disinfecting washers
- Disinfection washers for endoscopes
- Steam sterilizers
- Blood banks
- Medical refrigerators
- Laboratories
- Freezers
- Pharmaceutical industry:
- Steam Sterilizers
- Stabilization chambers
- Cooling rooms
- Storage rooms
- Incubators
- Laboratories
- Pharmaceutical freezers
- Food processing:
- Meat Packing Plants
- Pasteurization
- Spiral cooling/heating
- Transport equipment
- Storage freezing and cooling facilities
- Food smoking rooms
- Ovens
- Water autoclaves
The kit includes:
- 5x Ebro EBI 12-T441 Durable Temperature Logger
- 1x Ebro EBI 12-TP226 Rugged temperature and pressure logger
- EBI IF 200 Interface for remote data download from the loggers
- Silicone protection for EBI 12-TP226 logger
- Winlog.validation software with TÜV certification for data analysis
- Silicone probe holder set
- Carrying case
Features of the Ebro EBI 12-T441:
- Two external temperature sensors, axial, flexible, O 1.5mm, length 500mm
- Temperature measurement range -200...+200°C
- Temperature measurement accuracy:
- ±1.5°C (-200°C...-85°C)
- ±0.5 °C (-85 °C...-40 °C)
- ±0.2 °C (-40 °C...0 °C)
- ±0.1 °C (0 °C...+120 °C)
- ±0.05 °C (+120 °C...+140 °C)
- ±0.1 °C (+140 °C...+150 °C)
- ±0.5 °C (+150 °C...+200 °C)
- Temperature measurement resolution 0.01°C
- Memory for 2x 50,000 measurements
- Wireless data transmission
- IP68 / NEMA 6P
Features of the Ebro EBI 12-TP226:
- One external temperature sensor, axial, flexible, O 1.5 mm, length 250mm
- One internal pressure sensor, piezoresistive
- Temperature measurement range 0...+150°C
- Temperature measurement accuracy:
- ±0.1°C (0°C...+120°C)
- ±0.05 °C (+120 °C...+140 °C)
- ±0.1 °C (+140 °C...+150 °C)
- Temperature measurement resolution 0.01°C
- Pressure measurement range 1...4000 mbar
- Accuracy of pressure measurement:
- ±10 mbar (50...150mbar)
- ±10 mbar (2050...2250mbar)
- ±10 mbar (3000...3250mbar)
- ±15 mbar rest of range
- Pressure measurement resolution 1 mbar
- Memory for 2x 33,400 measured values
- Wireless remote data transmission
- IP68 / NEMA 6P